Epoxy Body Bulb Electrode – FLS pH/ORP 200

The FLS line of pH and ORP electrodes offers a cost-effective, multi-purpose solution for in-line or submersion measurements across a wide range of applications. Available in single and double junction versions, these electrodes can be supplied with or without quick-disconnect top caps. For automatic temperature compensation (ATC), a pH model with an integrated temperature sensor is also available.

Constructed with durable epoxy bodies, these electrodes deliver excellent chemical resistance, making them suitable for varied applications. Installation is simple—economic in-line mounting can be achieved with a reusable gland, while submersion mounting requires only a ½” or ¾” coupler with pipe extension. A special version is also designed for direct installation on FLS T-fittings or FLS clamp saddles with just the addition of a nut.

Epoxy Body Bulb Electrode – FLS pH/ORP 200

Key Features Of Epoxy Body Bulb Electrode – FLS pH/ORP 200

The FLS line of pH and ORP electrodes offers a cost-effective, multi-purpose solution for in-line or submersion measurements across a wide range of applications. Available in single and double junction versions, these electrodes can be supplied with or without quick-disconnect top caps. For automatic temperature compensation (ATC), a pH model with an integrated temperature sensor is also available.

Constructed with durable epoxy bodies, these electrodes deliver excellent chemical resistance, making them suitable for varied applications. Installation is simple—economic in-line mounting can be achieved with a reusable gland, while submersion mounting requires only a ½” or ¾” coupler with pipe extension. A special version is also designed for direct installation on FLS T-fittings or FLS clamp saddles with just the addition of a nut.

FLS pH_ORP 200 image

Technical Details of Epoxy Body Bulb Electrode – FLS pH/ORP 200

Specification Details
Operating Range – pH Electrodes: 0 – 14 pH (0 – 12.3 pH without Na+ error)
– ORP Electrodes: ± 2000 mV
Temperature Compensation Device (for TC model) PT1000
Pipe Size Range DN15 to DN100 (0.5” to 4”)
Zero Voltage Point New Electrode Performances 7.00pH ± 0.2pH
Efficiency New Electrode Performances > 97% @ 25°C (77°F)
Response Time New Electrode Performances – pH: 2 sec for 95% of signal change
– ORP: Application dependent
Reference Electrolyte Solidified gel 3.5M KCl for single junction versions KCl-KNO3 for double junction versions
Process Connection – In-line installation with threaded nipple ½”, ¾” or PG13,5
– FLS installation fittings – Submersible installation
Max Working Pressure / Working Temperature – 7 bar (100 psi) @ 25°C (77°F) – 1 bar (14.5 psi) @ 65°C (149°F)
Wetted Materials – Body: Epoxy – O-ring Junction: Silicone
– Junction: Pellon® – Sensing Surface: Glass membrane (pH), Platinum (ORP)
O-ring NBR (PH222 CD, PH223 CD, ORP222 CD, ORP223 CD)
Manufactured under ISO 9001, ISO 14001
Certifications CE, EAC
Technical Measurement of Epoxy Body Bulb Electrode – FLS pH/ORP 200
Epoxy Body Bulb Electrode - FLS pH_ORP 200 Drawing
Model Body Junction material/type Reference solution Sensing surface O-ring Connection Max working pressure
PH200C epoxy nylon/S.J. 3.5M KCl glass membrane - 5 mt. (16.5 ft.) Cable 7 bar @ 25°C / 1 bar @ 65°C (100 psi @ 77°F / 14.5 psi @ 149°F)
PH222CD epoxy nylon/D.J. 3.5M KCl / Sat’d KNO₃ glass membrane silicone Twist-Lock (BNC) same as above
PH223CD epoxy nylon/D.J. 3.5M KCl / Sat’d KNO₃ glass membrane silicone Twist-Lock (BNC) Same as above
ORP200C epoxy nylon/S.J. 3.5M KCl platinum - 5 mt. (16.5 ft.) Cable same as above
ORP222CD epoxy nylon/D.J. 3.5M KCl / Sat’d KNO₃ platinum silicone Twist-Lock (BNC) same as above
ORP223CD epoxy nylon/D.J. 3.5M KCl / Sat’d KNO₃ platinum silicone Twist-Lock (BNC) same as above
PH222CDTC epoxy nylon/D.J 3.5M KCl / Sat’d KNO₃ glass membrane - 5 mt. (16.5 ft.) Cable same as above

A → PH200C, PH222CDTC, ORP200C
B → PH222CD, ORP222CD
C → PH223CD, ORP223CD

1 → Cable: 5 mt (6.5 ft.)
2 → Epoxy body
3 → pH glass bulb
4 → BNC connector
5 → Buna-N O-rings
6 → FPM O-rings

Epoxy Body Bulb Electrode – FLS pH/ORP 200

Technical Specifications of Epoxy Body Bulb Electrode – FLS pH/ORP 200

  • Epoxy body
  • Single or double junction technology
  • Large gel reference volume
  • Easy and quick installation system
  • Outline cable or BNC connection
  • Version with temperature sensor combined
  • Special versions on request
  • Low cost fittings

Custom Product Manufacturing Available

If you are searching for customized product designs that meet your requirements, don’t worry. Petron Thermoplast also offers custom product manufacturing services to meet the unique needs of various industries. Work with our expert engineers to create custom designs suited for your application.